From Design to Performance: Navigating the Phases of Pharmaceutical Qualification

Introduction

Qualification is an important aspect of the pharmaceutical industry that ensures quality manufacturing processes and compliant end products. It is the process of demonstrating and documenting that equipment and ancillary systems are properly installed, work correctly, and actually lead to the expected results.

The main goals of qualification are to reduce risks, assure quality, and comply with regulations and standards. It provides objective evidence and assurance that a piece of equipment or system consistently performs its intended function according to predetermined specifications.

Qualification typically involves several key phases:

  • User Requirements Specification – The user requirements and expectations for functionality are documented.

  • Design Qualification – The proposed design of the equipment or system is reviewed to ensure it meets user requirements.

  • Installation Qualification – Installation of equipment is verified to comply with approved design specifications.

  • Operational Qualification – Tests are conducted to demonstrate the equipment operates as intended across specified operating ranges.

  • Performance Qualification – The equipment is evaluated to ensure it consistently performs according to predetermined acceptance criteria in real-world conditions.

Proper qualification is crucial for drug manufacturing facilities to prove their processes are robust and will reliably produce quality medicines. Executing each phase thoroughly is important to identify and mitigate risks prior to routine operation. This helps ensure patient safety through compliant pharmaceutical manufacturing.

User Requirement Specification (URS)

The User Requirement Specification (URS) is the first step in the qualification process. The purpose of the URS is to clearly define the functionality and operational requirements of the equipment or system. This document details what the equipment is intended to do and how it will be used in the manufacturing process.

The URS is developed jointly by the pharmaceutical company and the equipment manufacturer. The company provides details on their manufacturing process, product specifications, safety requirements, and any regulatory compliance needs. The manufacturer contributes their expertise on equipment capabilities and limitations.

Key details collected in the URS include:

  • Intended use and operating environment
  • Product contact materials and compatibility
  • Cycle times and throughput requirements
  • User interfaces, software, reporting needs
  • Alarm and interlock systems
  • Cleaning and maintenance requirements
  • Calibration and re-qualification frequency
  • Compliance with cGMP, regulations, standards

The URS review process validates that the company requirements align with the manufacturer’s equipment offerings. The final approved URS provides the basis for the equipment design and guides qualification activities moving forward.

Design Qualification

Design Qualification (DQ) is an important phase in pharmaceutical equipment qualification that ensures the proposed equipment design meets the specified user requirements and compliance standards.

The purpose of DQ is to verify that the equipment design is suitable for the intended use and will reliably perform required functions. It provides documented evidence that the equipment has been properly designed before proceeding with procurement and installation.

During DQ, the pharmaceutical company first prepares initial drawings and documentation outlining their specific needs and desired equipment capabilities. These user requirement specifications are then forwarded to the equipment manufacturer.

The manufacturer reviews the proposed design and may suggest modifications if needed to meet operability, cleanability, compliance, safety, or other requirements. The company and manufacturer collaborate iteratively on draft drawings and documentation until the design is finalized.

The DQ phase concludes once the company reviews the manufacturer’s proposed design and formally approves it. This signifies both parties agree the equipment design and supporting documentation conform to specified requirements prior to construction.

DQ ensures effective two-way communication occurs between the company and manufacturer early in the project. This helps avoid costly rework from design flaws being identified too late. Executed properly, DQ provides assurance that the final equipment will function as intended when installed at the company’s facility.

Site Acceptance Test (SAT)

The purpose of the Site Acceptance Test (SAT) is to verify that the equipment being purchased meets all requirements and specifications before it is shipped from the vendor’s site. This testing is performed by representatives from the pharmaceutical company at the equipment manufacturer’s facility.

During the SAT, company representatives will:

  • Visually inspect the equipment for any signs of damage or defects in workmanship. All components should be checked against the approved drawings and specifications.

  • Witness the operation of the equipment to ensure it functions as intended. The vendor should demonstrate all modes of operation and any safety interlocks or alarms.

  • Review calibration and test documentation provided by the manufacturer to confirm results meet expected tolerances and specifications.

  • Examine welds, wiring, cabling, fittings and other critical aspects to check for quality of work.

  • Confirm software meets functionality requirements by running simulated batch processes or test protocols. Any software bugs or issues should be documented.

  • Verify that all necessary spare parts, accessories, and documentation will be provided as agreed upon in the equipment contract.

The company representatives overseeing the SAT play a critical role in ensuring the purchased equipment is completely acceptable before releasing it for shipment. By thoroughly inspecting and testing the equipment at the vendor’s facility, costly problems or defects can be identified and corrected early in the project timeline. This provides assurance that the delivered equipment will meet Quality standards and avoid delays during installation and qualification at the pharmaceutical company’s facility.

Factory Acceptance Test

Factory Acceptance Test (FAT) is an important step in the pharmaceutical equipment qualification process. It is conducted after the equipment is manufactured and before it is shipped to the pharmaceutical facility.

The main purpose of FAT is to validate that the equipment meets the approved design specifications and performs as intended by the manufacturer. It provides assurance that the equipment functions properly before being installed on the pharmaceutical site.

During FAT, representatives from the pharmaceutical company visit the equipment manufacturer’s facility to witness and approve the testing. The following tests are typically performed:

  • Visual inspection to ensure the equipment is built according to approved drawings and specifications. Any deviations are documented.

  • Assembly check to verify correct manufacturing and assembly of all components.

  • Electrical safety tests to confirm the equipment meets safety standards.

  • Functional tests by running the equipment with placebo to confirm mechanical operations, process control capabilities, etc.

  • Factory control tests by integrating the equipment with its control system.

  • Simulation of worst-case process conditions to validate performance limits.

  • Software verification, if applicable.

  • Review of equipment manuals, certificates, documentation provided by vendor.

The equipment must pass all FAT protocols successfully as per acceptance criteria. If issues are identified, the manufacturer will need to fix and retest until equipment is approved.

Once FAT is completed successfully, the equipment will be cleared for shipment to the pharmaceutical facility. FAT provides assurance to the company that the equipment is ready for installation and subsequent qualification at site.

Installation Qualification

Installation Qualification (IQ) is performed after equipment is installed at the manufacturing site and before it is connected to other equipment. The purpose of IQ is to verify that the equipment is installed correctly according to the vendor’s recommendations and specifications.

Some of the checks performed during an IQ include:

  • Equipment location and access
  • Utilities like electricity, compressed air, nitrogen, etc. are connected and operating within specifications
  • All components, modules, and accessories listed on the purchase order have been installed
  • Calibration of measuring devices
  • Proper materials of construction
  • Installation per engineering drawings
  • Environmental conditions like temperature, humidity, and air changes within specifications
  • Safety systems in place
  • Proper signage and labeling
  • Cleaning and sanitization procedures defined

Thorough documentation is maintained for IQ including checklists, test records, inspection reports, notes, drawings, supplier manuals, etc. Deviations if any during the IQ are recorded, justification provided, and change order issued if required. Only after all installation qualification checks are cleared, is the equipment released for further qualification like Operational Qualification.

Operational Qualification

Operational qualification (OQ) involves testing equipment functions and operations against the operational specifications. The purpose of OQ is to verify that the equipment operates as intended across the full operating range.

Operational checks typically include:

  • Checking that the equipment operates properly at maximum and minimum operating limits for parameters like temperature, speed, pressure, voltage, etc. This confirms the equipment functions across its entire operating range.

  • Testing all modes, settings, safety features, alarms, interlocks, human-machine interfaces, etc. to confirm they operate to specifications. This verifies all operations and functions.

  • Confirming the equipment achieves the critical performance characteristics it was designed for, such as heating rate, flow rate accuracy, mixing speed, etc. This validates it meets operational requirements.

  • Running the equipment with actual or simulated products, ingredients, and components. This checks real-world conditions.

  • Testing integration with other equipment, systems, and facilities at the site. This verifies smooth interfacing.

  • Simulating worst-case scenarios like power failure to check backup systems or maximum load to check stress tolerance.

  • Checking ease of maintenance activities outlined in the maintenance plan.

The operational qualification provides evidence the equipment operates consistently and reliably as intended when put to use at the facility. This ensures it will function effectively for routine production.

Performance Qualification

Performance Qualification (PQ) is the process of demonstrating and documenting that equipment and ancillary systems perform as intended throughout all anticipated operating ranges.

The main purposes of PQ are:

  • To demonstrate that the equipment operates consistently and reproducibly within defined specifications and parameters for prolonged periods. This ensures that the equipment consistently produces the same results.

  • To verify that the equipment functions as intended over all anticipated operating ranges outlined in the User Requirements Specification (URS). The equipment is tested at minimum, normal, and maximum operating capacity levels.

  • To challenge equipment operating parameters through stress testing. This confirms that the equipment functions properly even under worst-case scenarios.

During PQ, detailed performance testing is conducted using actual raw materials, intermediates, and other production inputs. Equipment is run at maximum speed and capacity to verify performance under strenuous conditions. Multiple batches are produced to confirm reproducibility. Extensive laboratory testing is conducted to verify quality specifications are consistently met.

Maintenance

Once qualification of machinery is completed as per protocols, qualified status of equipment needs to be maintained by carrying out preventive maintenance activities as defined in equipment manuals provided by vendors. Typical maintenance activities include:

  • Regular cleaning and sanitization as per approved SOPs.
  • Periodic calibration of measuring devices like pressure gauges, temperature probes etc.
  • Timely replacement of parts nearing end of life like gaskets, pumps, motors.
  • Monitoring of equipment for any deviation from normal operation and troubleshooting.
  • Maintaining logs of all maintenance activities.

In addition to routine maintenance, requalification of equipment may be required under certain circumstances like:

  • Major changes done to equipment like addition of new features, changing critical parts.
  • Relocation of existing equipment to a new area.
  • Significant time period elapsed since last qualification.
  • Failure of equipment to meet product quality criteria.

The extent of requalification depends on the changes made and is determined by cross-functional teams. Typical requalification activities can involve partial or complete installation/operational/performance re-qualification.

Regular maintenance coupled with requalification ensures continued validated state of equipment over its lifetime. This provides ongoing assurance that the equipment consistently performs its intended functions to produce quality products.

Conclusion

Qualification is a vital process in the pharmaceutical industry to ensure that equipment and facilities meet the required standards for cGMP manufacturing. The several stages outlined – from user requirements, through design approvals, site and factory acceptance, and final installation, operational, and performance qualification – provide a rigorous framework to validate that everything functions as intended.

Strictly adhering to qualification protocols ensures that facilities, utilities, equipment, processes, and automation are carefully designed, installed, tested, and documented to comply with regulations. This provides confidence that drug products will consistently meet quality attributes and specifications. Any deviations or deficiencies can be identified and corrected.

Thorough qualification is not simply a regulatory exercise but an important investment that leads to right-first-time production, operational efficiency, product quality, and patient safety. The integrity of qualification processes directly impacts pharmaceutical quality systems. Following established guidelines and best practices for qualification ensures that manufacturing facilities and equipment perform reliably and consistently.

(Visited 51 times, 1 visits today)

Leave a Reply